X-Rite in Ergosoft RIP – Drivers, Connectivity, FAQ

X-Rite in Ergosoft RIP – Drivers, Connectivity and FAQ

Applies to: Ergosoft RIP 17 and later · X-Rite measuring devices (i1Pro, i1iO, i1iSis, eXact series, Ci6x, SP6x, Spectrolino / SpectroScan)

Last updated: Ergosoft RIP 17.8.2.0650 · 9 Nov 2026


Overview

Ergosoft RIP uses X-Rite spectrophotometers for all color measurements: Density Linearization, ColorGPS, ColorQPC, SwatchBook and SwatchMatcher. You select and detect the device once in Tools > Application Defaults > Application > Spectrophotometer. The measurement standards set there apply to all measurements.

DeviceTypeNotes
X-Rite i1Pro, i1Pro 2, i1Pro 3, i1Pro 3 PlusHandheld spectrophotometerScan and spot measurement. The Plus series has a large aperture and supports M3.
X-Rite i1iO, i1iO 2, i1iO 3Scanning table for the i1ProAutomatic chart measurement, spot and scan mode
X-Rite i1iSis, i1iSis 2Automatic chart readerSheet charts
X-Rite eXact, eXact 2 seriesHandheld spectrophotometerSpot (and scan) measurement
X-Rite Ci62, Ci64, Ci64UV, SP6x familyHandheld sphere spectrophotometersSpot measurement
X-Rite Spectrolino / SpectroScanLegacy spectrophotometer and tableSerial port or COM to USB adapter

For the current list, see Supported Measurement Devices on the Ergosoft website. The X-Rite device drivers are included in the RIP installer; you do not need an extension or Updater. The USB hardware driver of X-Rite is installed separately (see Installation and Drivers).


Latest Driver Updates

BuildModel / driverChange
17.8.2.0650X-Rite eXact 2 seriesSDK update (v1.5.3)
17.7.2.0465All measuring devices (Application Defaults)New CIE Illuminant/Observer subvariants, selectable chromatic adaptation (CAT) methods, and the density standard SCTV (ISO 20654)

Ergosoft RIP 17.7.2.0465 has no X-Rite-specific driver change. Older changes: see the public changelog (category Driver, search "X-Rite").


Measuring Devices

Installation and Drivers

Which X-Rite devices can I use with Ergosoft RIP?
The devices in the overview table: i1Pro, i1iO and i1iSis series (all generations), eXact and eXact 2 series, Ci62 / Ci64 / Ci64UV, SP6x family and the legacy Spectrolino / SpectroScan. The first generations (i1Pro, i1iO, eXact) are still supported. The eXact 2 series is supported since Ergosoft RIP 17.1.2.8882. Applies to: all versions.

Do I need to install a driver for my X-Rite device?
The X-Rite device drivers are part of Ergosoft RIP. The USB hardware driver of the device comes from X-Rite: install it separately and check in the Windows Device Manager that Windows recognizes the device without errors. Applies to: all versions.

Ergosoft RIP cannot connect to the device, or the connection drops. What do I check?
Close all other applications that can use the device, for example X-Rite software or other color software. If another program accesses the device, it blocks the communication with Ergosoft RIP. Then detect the device again. Applies to: all versions.

How do I select and detect my X-Rite device?
Go to Tools > Application Defaults > Application > Spectrophotometer, select the device under Type and click Detect/Set. Ergosoft RIP shows the device name and serial number. Options opens the settings of the connected device. In Manage Measurements, the Spectrophotometer button opens the detection again, for example to reconnect the device. Applies to: all versions.

Which device type do I select for an i1Pro or i1Pro 2?
Select X-Rite i1Pro 1/2: the first and second generation share one entry in the Type list. The i1Pro 3 series has its own entries. Applies to: all versions.

Can I use X-Rite sphere devices on a ColorScout A+ table?
Yes. The Ci62, Ci64, Ci64UV and SP6x family are also supported on the ColorScout A+ scanning table from ColorPartner. A device on a ColorScout A+ table works very much like an i1Pro on the X-Rite i1iO table. See ColorScout A+ (ColorPartner). Applies to: all versions.

Can I still use an X-Rite Spectrolino / SpectroScan?
Yes, also with a COM to USB adapter. Ergosoft RIP 17.3.2.9232 fixed a communication problem with these adapters. Applies to: Ergosoft RIP 17.3.2.9232 and later.

Connection and Calibration

My X-Rite i1 does not calibrate. What do I check?
Open the cover of the calibration base, so that the white tile is not covered. Clean the white tile and check that the device sits fully on the base. If calibration still fails, the device may need service. See X-Rite i1 - Will not calibrate. Applies to: all versions.

Why does the RIP ask for a white calibration?
The device must calibrate on its white tile before it measures. Follow the prompt, then measure the chart. Applies to: all versions.

My i1Pro 3 Plus is not recognized as a Plus device, and M3 is missing. What do I do?
A specific sequence of user actions could make the RIP lose the Plus detection, which hides M3. Update to Ergosoft RIP 17.5.2.0098 or later. Applies to: Ergosoft RIP 17.5.2.0098 and later.

Measurement Modes and Standards

Where do I set the measurement condition (M0, M1, M2, M3) and the other device options?
Click Options next to Detect/Set in Tools > Application Defaults > Application > Spectrophotometer. The dialog shows only the features of the connected device, for example the Measurement Mode, the Illumination Condition (M0, M1, M2 and, on the i1Pro 3 Plus, M3) and the Patch Recognition. The measuring wizard, which all measurements use (Density Linearization, ColorGPS, SwatchBook, ColorQPC), shows the measuring conditions again at the start of each session, so you can check and change them. Use the same condition for all measurements of one workflow (linearization, profile and verification). M3 is hidden for the i1Pro 3 regular aperture since Ergosoft RIP 17.2.2.9069. Applies to: all versions.

Which measurement method does the RIP use for the i1Pro 3 by default?
The defaults follow the X-Rite recommendations. Change them in Application Defaults only if your workflow requires it. Applies to: all versions.

Which density standard do I use?
Set it under Measurement Standards in the same dialog, separately for CMY inks, black ink and additional inks. The default is Status E - ISO 5-3. Status T (common in the United States), Status I and DIN 16536-2 (legacy) are also available. For additional inks, the default is Ergosoft Dynamic Filter; the SCTV method (ISO 20654) is available since Ergosoft RIP 17.7.2.0465. Applies to: all versions.

Why do my density values differ from an older RIP version?
The default density standard is Status E - ISO 5-3. To get values that match the older DIN standard, select DIN 16536-2 (Legacy) in the measurement standards. Applies to: all versions.

Do I need to change the illuminant, observer or chromatic adaptation?
No, for most workflows the default (D50, 2°, XYZ white point scaling) is correct. Since Ergosoft RIP 17.7.2.0465, you can also select the subvariants Packaging (D50, 10°) and Digital Imaging (D65, 2°) and other CAT methods. Change these settings only if your color workflow requires it. Applies to: Ergosoft RIP 17.7.2.0465 and later.

Charts and Scanning Tables

Which chart geometry do I select?
Select the geometry with the name of your device series. The i1iO 3 and i1Pro 3 have dedicated geometries for spot and scan mode. Do not use a 3 series geometry with an i1Pro 2 or i1iO 2. Applies to: Ergosoft RIP 17.1.2.8882 and later.

Can I measure in scan mode with the i1iO 3?
Yes. The i1iO 3 (Standard and Plus) offers an optional scan measurement method. Select Scan measurement instead of Single Measurement in the measurement wizard. Applies to: Ergosoft RIP 17.1.2.8882 and later.

Can I measure single patches with the i1Pro 3?
Yes. The i1Pro 3 (Plus) supports spot reading in addition to scanning. Applies to: all versions.

Can I measure in scan mode with the eXact or eXact 2?
The eXact devices also have a scan mode, but the results are less reliable than with spot measurement. Use spot measurement for charts. Applies to: all versions.

The RIP reports a large difference between repeated measurements. What do I do?
Select one of the options in the prompt: keep the previous measurements, use the last measurement, or average them (not recommended in normal cases). With 3 or more measurements per patch, the wizard compares each reading with the average and Delete Marked removes outliers. Applies to: Ergosoft RIP 17.6.2.0282 and later.

Manage Measurements with the Spectrophotometer button

Where can I find my measurements again?
After each successful measurement session, the RIP saves the data automatically in [UserDataFolder]\Data\Color\Measurements\AutoSave as a .cxf file. Applies to: Ergosoft RIP 17.5.2.0098 and later.

Can I use an X-Rite i1Pro 3 with HP Easy Profiling?
No. HP Easy Profiling does not support the i1Pro 3. Use a device that HP Easy Profiling supports, or use Density Linearization and ColorGPS in the RIP. See X-Rite i1 Pro 3 not working with HP Easy Profiling. Applies to: all versions.

My X-Rite device is not in the list. Can I still use it?
Yes, if the device software can export spectral values as an ASCII file in the required format. Select From File as the device and load the file. Density-only data works for Density Linearization, but not for ColorGPS. See Density Linearization. Applies to: all versions.


Known Issues

No open known issues for X-Rite devices in Ergosoft RIP 17.8.2.0650.


References

  • Changelog #13161 – Ergosoft RIP 17.1.2.8882: X-Rite eXact: an issue with multiple, sequential device initializations has been resolved.
  • Changelog #13191 – Ergosoft RIP 17.1.2.8882: X-Rite i1iO 3 (Standard and Plus series) has been updated to the latest SDK; the optional scan measurement method has been added.
  • Changelog #13276 – Ergosoft RIP 17.1.2.8882: New measurement chart geometries for X-Rite i1Pro 3 (Standard and Plus) for modes spot and scan have been added.
  • Changelog #13366 – Ergosoft RIP 17.1.2.8882: X-Rite eXact 2 series (new measuring device).
  • Changelog #14172 – Ergosoft RIP 17.2.2.9069: X-Rite i1Pro 3 (regular aperture): M3 is generally not supported by this device and is now hidden.
  • Changelog #14616 – Ergosoft RIP 17.3.2.9232: X-Rite Spectrolino / Spectroscan: Fixed a communication problem with the device using a COM to USB adapter.
  • Changelog #15274 – Ergosoft RIP 17.5.2.0098: X-Rite i1Pro 3 Plus (large aperture): A specific sequence of user interactions could cause the device to no longer be recognized as a Plus series.
  • Changelog #15301 – Ergosoft RIP 17.5.2.0098: Color measurement data is automatically saved after each successful measurement session (AutoSave).
  • Changelog #15935 – Ergosoft RIP 17.7.2.0465: New global setting for chromatic adaptation (CAT) methods under Application Defaults > Spectrophotometer > Measurement Standards.
  • Changelog #3505 – Ergosoft RIP 17.7.2.0465: The Spot Color Tone Value (SCTV - ISO 20654:2017) method is available as density standard.
  • Changelog #16061 – Ergosoft RIP 17.8.2.0650: X-Rite eXact 2 Series (v1.5.3) SDK update.